Name: AAA patients
Description: Protein expression in the Inferior mesenteric vein (IMV) of patients with AAAs
Version: MIAPE: Gel Electrophoresis 1.4
1. General features
1.1.1 Date Stamp
2009-03-01
1.1.2 Responsible person or role
Affiliation: Researcher
(i) Name: Ian Nordon
(ii) Postal address: St George's Vascular Institute
(iii) Email address: inordon@sgul.ac.uk
1.1.3 Electrophoresis type
Difference gel electrophoresis
2. Sample
2.1.1 Sample Name(s)
-
- Sample name: AAA patients
- Sample type: Test sample
2.1.2 Loading buffer
- Urea (Amersham Biosciences)8M, 48g Pharmalyte (Amersham Biosciences)0.2%, 2.5(ml) CHAPS (VWR International), 0.5%, 0.5g DTT (Sigma), 2%, 0.4g In 100ml double distilled water. Store at -80ºC
3. Gel matrix and electrophoresis protocol
3.1 Dimension details
3.1.1 Ordinal number for this dimension
First
3.1.2 Separation method employed
Isoelectric focusing (IEF)
3.2 Gel Matrix
3.2.1 Description of gel matrix
IPG strip
Denaturing
3.2.2 Gel manufacture
Gel was manufactured using the following reference protocol:
Citation:
Industry standrard,n/a,
page(s) n/a (2003).
URL:
http://www.signaling-gateway.org/data/cgi-bin/ProtocolFile.cgi/afcs_PP00000149.pdf?pid=PP00000149
3.2.3 Physical dimensions
X: 180 mm
Y: 3 mm
Z: 0.5 mm
3.2.4 Physiochemical property range and distribution
linear pH 3 - 10
3.2.5 Acrylamide concentration
8-12 % (linear)
3.2.6 Acrylamide : Crosslinker ratio
Crosslinker: Bisacrylamide
Ratio: 19:1
3.2.7 Additional substances in gel
APS 38ml
Glycerol 149ml
Temed 44ml
Tris-SDS 151ml
Water 686ml
3.2.8 Gel lane
13.2.9 Sample application
Lane 1
- Sample: AAA patients
- Volume of sample: 50 µL
- Loading buffer: Urea (Amersham Biosciences)8M, 48g Pharmalyte (Amersham Biosciences)0.2%, 2.5(ml) CHAPS (VWR International), 0.5%, 0.5g DTT (Sigma), 2%, 0.4g In 100ml double distilled water. Store at -80ºC
- Volume of loading buffer: 320 µL
Loading method: rehydration loading.
3.3 Protocol
3.3.1 Buffers
Tris-Glycine SDS Running Buffer
3.3.2 Electrophoresis conditions
Running temperature: 10 °C
Hold: 500 V, 10 min
Hold: 500 V, 110 min
Hold: 1200 V, 45 min
Hold: 2000 V, 30 min
Hold: 2500 V, 20 min
Hold: 3500 V, 20 min
Hold: 3500 V, 10 min
Hold: 10000 V, 180 min
Hold: 10000 V, 60 min
Hold: 500 V, 60 min
4. Inter-dimension Process
4.1 Protocol
4.1.1 Step name
equilibration
4.1.2 Inter dimension buffer
Equilibration buffer
Urea Amersham Biosciences 8M 48
SDS Invitrogen 2% 20
Glycerol Invitrogen - 300g
Tris pH 8.8 Invitrogen 50mM 33(ml)
4.1.3 Additional reagents
No additional reagent.
4.1.4 Equipment
4.1.5 Protocol
3. Gel matrix and electrophoresis protocol
3.1 Dimension details
3.1.1 Ordinal number for this dimension
Second
3.1.2 Separation method employed
Sodium dodecyl sulphate polyacrylamide gel electrophoresis (SDS-PAGE)
3.2 Gel Matrix
3.2.1 Description of gel matrix
slab gel
Denaturing
3.2.2 Gel manufacture
Gel was manufactured using the following reference protocol:
Citation:
Industry standrard,n/a,
page(s) n/a (2003).
URL:
http://www.signaling-gateway.org/data/cgi-bin/ProtocolFile.cgi/afcs_PP00000149.pdf?pid=PP00000149
3.2.3 Physical dimensions
X: 260 mm
Y: 210 mm
Z: 2 mm
3.2.4 Physiochemical property range and distribution
linear apparent molecular mass 200 - 10 kDa
3.2.5 Acrylamide concentration
8-12 % (linear)
3.2.6 Acrylamide : Crosslinker ratio
Crosslinker: Bisacrylamide
Ratio: 19:1
3.2.7 Additional substances in gel
APS 38ml
Glycerol 149ml
Temed 44ml
Tris-SDS 151ml
Water 686ml
3.2.8 Gel lane
13.2.9 Sample application
Loading method: well loading.
3.3 Protocol
3.3.1 Buffers
Tris-Glycine SDS Running Buffer
3.3.2 Electrophoresis conditions
Running temperature: room temperature (no cooling device).
Hold: 600 V, 18.5 h
5. Detection
5.1 Direct detection
5.1.1 Name of direct detection_process
Fluorescent staining
5.1.2 Direct detection agents
Cyanine dyes (Cy2/3/5)
5.1.3 Additional reagents and buffers
No additional reagents or buffer
5.1.4 Equipment
Manufacturer: GE Healthcare
Model: Typhoon
Model number: 9400
5.1.5 Direct detection protocol
Detection is described in the following reference protocol:
Citation:
Other,n/a,
page(s) n/a (2003).
URL:
http://www.currentprotocols.com/protocol/mb1023
6. Image Acquisition
6.1 Acquisition Equipment
6.1.1 Type of equipment
laser scanner6.1.2 Name of equipment
Manufacturer: GE Healthcare
Model: Typhoon
Model number: 9400
6.1.3 Software
Manufacturer: GE Healthcare
Model: ImageQuant
Model number: 5.2
6.1.4 Calibration
Yes (automatic)
6.1.5 Equipment specific parameters
Default (vendor) parameters.6.2 Acquisition Protocol
6.2.1 Image acquisition process
30 minutes/gel100 micrometer pixel size
Cy2 500nM
Cy3 545nM
Cy5 550nm
6.2.2 Reference to gel matrix
There is only one gel in this document.
7. Image
7.1.1 Image name (or id)
AAA (format: TIFF)
7.1.2 Dimensions
Width: 2800 px
Height: 2080 px
7.1.3 Resolution
100 µm/px
7.1.4 Bit-depth
16-bit (HighColor)7.1.5 Image location
E:\AAA1.tif7.1.6 Standard image orientation
Yes